BCA / Bradford Standard Curve Calculator

Fit a linear standard curve to your standards and calculate sample concentrations from their absorbance, with R².

A = m × C + b; C(sample) = (A − b) ÷ m × dilution

Fill in the fields above — results update live.

Computed locally — no data leaves your browser. Check results against your protocol and reagent datasheets before use.

BCA and Bradford curves often curve slightly at the top end. If R² is low, drop the highest standard or use a quadratic / 4PL fit.

Worked example

Inputs: Standards — one per line: concentration, absorbance 0, 0.052 ; 125, 0.178 ; 250, 0.301 ; 500, 0.552 ; 750, 0.790 ; 1000, 1.03; Sample absorbance(s) — separate with commas or spaces 0.412, 0.655; Sample dilution factor 1.

  • Fit: A = 9.781e-4 × C + 0.05591
  • R²: 0.9999
  • Sample 1 (A = 0.412): 364.1
  • Sample 2 (A = 0.655): 612.5

Common questions

What if my standard curve is not linear?

BCA and Bradford curves often flatten at high concentrations. Drop the top standard or use a quadratic or 4PL fit.

How should I handle the blank?

Include a zero-concentration standard, or subtract the blank absorbance from every reading before entering the values.