Calculators · Protein
BCA / Bradford Standard Curve Calculator
Fit a linear standard curve to your standards and calculate sample concentrations from their absorbance, with R².
A = m × C + b; C(sample) = (A − b) ÷ m × dilution
Fill in the fields above — results update live.
Computed locally — no data leaves your browser. Check results against your protocol and reagent datasheets before use.
BCA and Bradford curves often curve slightly at the top end. If R² is low, drop the highest standard or use a quadratic / 4PL fit.
Worked example
Inputs: Standards — one per line: concentration, absorbance 0, 0.052 ; 125, 0.178 ; 250, 0.301 ; 500, 0.552 ; 750, 0.790 ; 1000, 1.03; Sample absorbance(s) — separate with commas or spaces 0.412, 0.655; Sample dilution factor 1.
- Fit: A = 9.781e-4 × C + 0.05591
- R²: 0.9999
- Sample 1 (A = 0.412): 364.1
- Sample 2 (A = 0.655): 612.5
Common questions
What if my standard curve is not linear?
BCA and Bradford curves often flatten at high concentrations. Drop the top standard or use a quadratic or 4PL fit.
How should I handle the blank?
Include a zero-concentration standard, or subtract the blank absorbance from every reading before entering the values.